A silicon-on-insulator diode array with a sensitive depth of 10 microns has been developed for microdosimetry in proton therapy. The detector was coupled to a radiation-hard charge sensitive amplifier with the probe assembly capable of measuring an LET down to 1.2 keV/ym. The device has been successfully tested at two proton therapy centers: The 230 MeV Northeastern Proton Therapy Center, Boston and the 250 MeV Proton Medical Research Center at Tsukuba, Japan. The device offers much improved spatial resolution compared with a proportional gas counter particularly in the critical high dose region around the proton Bragg peak. Due to its small cross-sectional area (0.04 cm2) measurements may also be made in facilities with short high intensity beams.
Abstract-Silicon microdosimetry measurements in Fast Neutron Therapy were simulated using the GEANT4 Monte Carlo toolkit. The possibility of using silicon microdosimeters for verification of Monte Carlo based treatment planning systems in hadron therapy is suggested.Index Terms-Fast neutron therapy (FNT), microdosimetry, Monte Carlo.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.