A preliminary assessment of the occupational dose to the intervention radiologist received in fluoroscopy computerised tomography (CT) used to guide the collection of lung and bone biopsies is presented. The main aim of this work was to evaluate the capability of the reading system as well as of the available whole-body (WB) and extremity dosemeters used in routine monthly monitoring periods to measure per procedure dose values. The intervention radiologist was allocated 10 WB detectors (LiF: Mg, Ti, TLD-100) placed at chest and abdomen levels above and below the lead apron, and at both right and left arms, knees and feet. A special glove was developed with casings for the insertion of 11 extremity detectors (LiF:Mg, Cu, P, TLD-100H) for the identification of the most highly exposed fingers. The H(p)(10) dose values received above the lead apron (ranged 0.20-0.02 mSv) depend mainly on the duration of the examination and on the placement of physician relative to the beam, while values below the apron are relatively low. The left arm seems to receive a higher dose value. H(p)(0.07) values to the hand (ranged 36.30-0.06 mSv) show that the index, middle and ring fingers are the most highly exposed. In this study, the wrist dose was negligible compared with the finger dose. These results are preliminary and further studies are needed to better characterise the dose assessment in CT fluoroscopy.
Computed tomography fluoroscopy (CTF) is a useful imaging technique to guide biopsies, particularly lung biopsies, but it also has the potential for very high hand exposures, despite use of quick‐check method and needle holders whenever feasible. Therefore, reliable monitoring is crucial to ensure the safe use of CTF. This is a challenge, because ring dosimeters monitor exposure only at the base of one finger, while the fingertips may be exposed to the highly collimated CT beam. In this work we have explored the possibility of using Gafchromic XR‐QA2 self‐developing film as a complementary dosimeter to quantify hand exposure during CTF‐guided biopsies. A glove used in a previous study and designed to contain 11 TLDs was adapted to include Gafchromic strips 7 mm wide, covering the fingers. A total of 22 biopsies were successfully performed wearing this GafTLD glove under sterile gloves, and the IR reported no difficulty or reduction of dexterity while wearing it. Comparison of dose distributions obtained from digitization of the Gafchromic film strips and absolute Hp(0.07) readings from TLDs showed good agreement, despite some positional uncertainty due to relative movement. Per procedure, doses at the base of the ring finger can be as low as 3%–8% of hand dose maximum. Accumulated dose at the base of the ring finger was four times lower than the dose maximum.PACS numbers: 07.57.Kp, 29.40.‐n, 85.25.Pb, 87.57.qp
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.