A low-field medium-resolution NMR spectrometer, with an operating frequency of 29 MHz for 1H, has been assessed for on-line process analysis. A flow cell that incorporates a pre-magnetisation region has been developed to minimise the decrease in the signal owing to incomplete polarisation effects. The homogeneous esterification reaction of crotonic acid and 2-butanol was monitored using a simple sampling loop; it was possible to monitor the progression of the reaction through changes in CH signal areas of butanol and butyl crotonate. On-line analysis of heterogeneous water-toluene mixtures proved more challenging and a fast sampling loop system was devised for use with a 5 L reactor. The fast sampling loop operated at a flow rate of 8 L min(-1) and a secondary sampling loop was used to pass a sub-sample through the NMR analyser at a slower (mL min(-1)) rate. It was shown that even with super-isokinetic sampling conditions, unrepresentative sampling could occur owing to inadequate mixing in the reactor. However, it was still possible to relate the 1H NMR signal obtained at a flow rate of 60 mL min(-1) to the composition of the reactor contents.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.