In this study, a glassy carbon electrode (GC) was modified with an electropolymerized film of 1-naphthylamine (1-NAP) with a subsequent overoxidation treatment in 0.2 M sodium hydroxide solution. This polymer p-1-NAPox film coated GC electrode was used for the selective determination of dopamine (DA) in the presence of a triple concentration of ascorbic acid (AA).These studies were performed using cyclic voltammetry and square-wave voltammetry at physiological pH. p-1-NAPox shows an attractive permselectivity, a marked enhancement of the current response and antifouling properties when compared to a bare GC electrode activated in basic media. With a preconcentration time of 3 minutes at open circuit, linear calibration plots were obtained for DA in buffer solution (pH 7.4) over the concentration range from 1 Â 10 À6 -1 Â 10 À4 M with a detection limit of 1.59 Â 10 À7 M.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.