The Abbott TDx fluorescence polarization immunoassay (FPIA) system has been evaluated and compared with well-established high performance liquid chromatography (HPLC) for the determination of three anticonvulsant drugs: phenytoin, phenobarbitone and carbamazepine. These assays were evaluated for precision, calibration curve stability, specificity and accuracy. Within-run precision studies using control samples (n = 15) in the subtherapeutic, therapeutic, and toxic concentrations, resulted in coefficients of variation in the range of 1.79-3.99% (FPIA) and 1.16-2.52% (HPLC), respectively. Between-run precision ranged from 2.32-6.34% for FPIA and from 2.04-3.38% for HPLC. Comparison of 122 patient samples assayed with both methods indicated an extremely good analytical correlation (r = 0.96) for all three comparisons. The FPIA method offers significant advantages in calibration curve stability while maintaining accuracy and precision comparable with those of established HPLC procedures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.