A hydrophilic interaction chromatographic (HILIC) procedure for the quantification of Sodium Cromoglicate (SCG) in ophthalmic solution is developed. Mobile phase consists of ACN and buffer, 86:14 v/v. Atlantis HILIC-Si column, 25 cm x 4.6 mm, is used as stationary phase. Detection is carried out using a variable wavelength UV-Vis detector at 326 nm. Linearity range and percent recoveries for SCG were 50-400 mug/mL and 100.44%, respectively. The SCG HILIC-UV assay was validated according to the International Conference on Harmonization guidelines. The method separates two impurities and degradation products resulting from stress environment. Influence of organic solvent, ionic strength and mobile phase pH on the retention of SCG is studied. The paper provides optimization of polar anionic solute (SCG) on unmodified silica by HILIC. Proposed method can be used as a stability-indicating assay for SGC and can be proved to be beneficial in ESI-MS for enhanced sensitivity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.