Indigenous instrumentation has been designed and developed for the measurement of the concentration of analytes from eight conductometric metal oxide sensors. The hardware scheme of instrumentation is based on the astable multivibrator configuration. The hardware measures the resistance output from the sensors, conditions, processes, and displays the data on the liquid crystal display. An 8051 based processor averages the data, converts them into engineering units, and sends them to remote PC through ethernet communication for post-data analysis. A graphical user interface (GUI) is developed to acquire, monitor, and display the eight channels’ sensor output. GUI plots the online data and offline data as a popup window. The hardware and software of the instrument were tested with standard resistors for calibration and found that in-house developed instrumentation is able to measure with an accuracy of ±0.5% with a resolution of 500 Ω. The instrument has been tested with a semiconductor metal oxide sensor, viz., chromium niobate (CrNbO4).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.