Ab st ra c t -There is a growing interest in accurately measuring the timing of the J peak of the ballistocardiogram (BCG) in order to obtain cardiovascular function markers non-invasively, especially in modern home healthcare applications. In this paper we have studied the effect that some common uncertainty sources have in the time measurement of the J peak. This is a necessary step towards the standardization of modern ballistocardiography systems equivalent to that available for EGC systems. We conclude that, to reduce J peak time uncertainty below the measured intrinsic uncertainty of about ±2 ms, the minimal bandwidth should be from 1.5 Hz to 22.5 Hz; the sampling frequency can be decreased up to 50 Hz when using cubic spline interpolation; 5 bits are required to quantify the signal, and signal-to-noise ratio (SNR) and signal-to-interference ratio (SIR) should be over 40 dB and 3 dB respectively. Ke ywo rd s: Ballistocardiogram, acquisition system, biomedical instrumentation, standardization.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.