This article describes a method that uses Carbon Screen‐printed Electrodes (C‐SPEs) to detect 3,4 – methylenedioxyamphetamine (MDMA) by Linear Sweep Voltammetry in aqueous medium. Major parameters of this technique were evaluated aiming improve the method sensibility. Amines interference were conducted in order to verify disturbs at the MDMA response. The method obtained a linear response from 1×10−5 mol L−1 to 1×10−4 mol L−1 with linear correlation coefficient of 0.996, Amperometric Sensitivity (AS) of 0.025 A×mol−1 L, Limit of Detection (LOD) of 1,83×10−6 mol L−1, and Limit of Quantification (LOQ) of 6,11×10−6 mol L−1. The method applicability, reproducibility and reproducibility were carried over inter/intra days tests and its application on seized samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.