Abstract. The article shows the results of calculations necessary for testing to approve the space qualified electric field probe used in outer space. A special electric field probe calibration setup was developed and manufactured, which allows testing a space qualified electric field probe in the entire measurement range (from ± 0.1 to ± 200 kV/m) under low (-80 °C) and high (+72 °C) temperatures and at reduced pressure of 1.3 · 10 -3 Pa (10 -5 torr). Metrological tests of the probes batch, calibrated by the space qualified electric field probe calibrator, performed at the verification equipment under normal conditions, showed that the maximum error of the electric field probe does not exceed 15% in the entire measurements range which will increase the accuracy of determining the satellite surface charging parameters and help in designing modern spacecrafts.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.