17Three quantification methodologies, namely calibration with internal standard (Cal-IS, non-weighted), 18 weighted calibration with internal standard (wCal-IS) and isotope pattern deconvolution (IPD) have been 19 used for the determination of testosterone in urine by LC-MS/MS. Uncertainty has been calculated and 20 compared for the three methodologies through intra-and inter-laboratory reproducibility assays. IPD 21showed the best performance for the intra-laboratory reproducibility, with RSD and combined uncertainty 22 values below 4% and 9% respectively. wCal-IS showed similar performance, while Cal-IS where not constant 23 and clearly worse at the lowest concentration assayed (2 ng/mL) reaching RSD values up to 16%. The inter-24 laboratory assay indicated similar results although wCal-IS RSD (20%) was higher than IPD (10%) and Cal-IS 25 get worse with RSD higher than 40% for the lowest concentration level. Uncertainty budgets calculated for 26 the three procedures revealed that intercept and slope were the most important factors contributing to 27 uncertainty for Cal-IS. The main factors for wCal-IS and IPD were the volumes of sample and/or standard 28 measured. 29
30
INTRODUCTION
31
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.