To develop simple electrochemical immunoassays, a screen printed amperometric microcell with graphite working and Ag/AgCl reference electrodes was tested for the determination of alkaline phosphatase enzyme (ALP) and anti-humanIgG conjugated ALP (alpha-hIgG-ALP) activity in 5-10 microl samples. To ensure reproducible, steady state conditions, the working electrode surface was coated with mass-transport controlling hydrogel layer. The kinetic response curves of the hydrogel coated electrodes were linear. In addition, the hydrogel layer reduced the nonspecific adsorption of the alpha-hIgG-ALP conjugate on the working electrode surface. The measurements were made in the range of 2 divided by 4000 mU ml(-1) enzyme activities using ascorbic acid 2-phosphate (AAP) as the enzyme substrate. AAP is commercially available, non-toxic and has excellent stability. The sensitivity of the determinations was about 71% of the sensitivity which could be achieved using p-aminophenylphosphate (PAPP), a not easily accessible and unstable enzyme substrate. The experimentally determined kinetic parameters of the ALP enzyme catalyzed reactions were the same with the bare and hydrogel layer coated electrodes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.