A new system for on-line preconcentration of molybdenum by sorption on a minicolumn associated to inductively coupled plasma -optical emission spectrometry with ultrasonic nebulization was studied. It is based on the sorption of molybdenum on a column packed with immobilized baker's yeasts on controlled pore glass without further complexing reagent. The molybdenum preconcentrated by biosorption was subsequently eluted with hydrochloric acid. Considering a sample flow rate of 5.0 mL min -1 , 10 mL of sample was preconcentrated in 2 min. achieving a sensitive total enhancement factor of 480-fold, and t he detection limit (3s) obtained was 21 ng L -1 .Additionally, the calculated precisions expressed as percent relative standard deviation (RSD%) was 1.9%.Satisfactory results were obtained for the determination of molybdenum in standard reference material NIST 1643e Trace Elements in Water and real water samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.