In the light of continuous improvement and optimization, recent experiments that the authors conducted give new insights into the applied evaluation method of Riegel et al. [1]: Thorough investigations of the previous results regarding the usage of the Lowry Assay showed discrepancies in the determination of the released amount of protein in the analysis solution. The accurate quantification of this parameter is crucial as it directly influences the calculation of the residual enzymatic activity. In concrete terms, this finding has a major impact on the presented and discussed results in the article “Activity determination of FAD‐dependent glucose dehydrogenase immobilized in PEDOT: PSS‐PVA composite films for biosensor applications” [1]. Thus, this commentary addresses the new insights concerning the applied evaluation method, explains necessary revisions and discusses new conclusions derived from the adjusted evaluation method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.