Objectives
The objective of our study is to evaluate the effect of storage temperature and time to analysis on arterial blood gas parameters in order to extend the CLSI recommendations.
Methods
Stability of 12 parameters (pH, pCO₂, pO₂, Na+, K+, Ca2+, glucose, lactate, hemoglobin, oxyhemoglobin, carboxyhemoglobin, methemoglobin) measured by GEM PREMIER™ 5000 blood gas analyzer was studied at room temperature and at +4 °C (52 patients). The storage times were 30, 45, 60, 90 and 120 min. Stability was evaluated on the difference from baseline, the difference from the analyte-specific measurement uncertainty applied to the baseline value, and the impact of the variation on the clinical interpretation.
Results
At room temperature, all parameters except the lactate remained stable for at least 60 min. A statistically significant difference was observed for pH at T45 and T60 and for pCO2 at T60 without modification of clinical interpretation. For lactate, clinical interpretation was modified from T45 and values were outside the range of acceptability defined by the measurement uncertainty. All parameters except pO2 remained stable for at least 120 min at +4 °C.
Conclusions
A one-hour transport at room temperature is compatible with the performance of all the analyses studied except lactate. If the delay exceeds 30 min, the sample should be placed at +4 °C for lactate measurement. If the samples are stored in ice, it is important to note that the pO2 cannot be interpreted.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.