Material and methods IVD was performed for 40 treatments planned using intra-operative trans-rectal ultrasound (TRUS) with a MOSFET inserted into an additional needle. Post-treatment TRUS images were acquired for 20 patients to assess needle movement. Monte Carlo simulations of treatment plans were performed for 10 patients to assess impact of heterogeneities.Per-needle and total plan uncertainties were estimated and retrospectively applied to the measured data as error detection thresholds.
ResultsThe mean measured dose was -6.4% compared to prediction (range +5.1% to -15.2%). Needle movement and heterogeneities accounted for -1.8% and -1.6% of this difference respectively (mean values for the patients analysed). Total plan uncertainty (k=2) ranged from 11% to 17% and per needle uncertainty (k=2) ranged from 18% to 110% (mean 31%). One out of 40 plans and 5% of needles were outside k=2 error detection threshold.
Conclusions
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.