We evaluated the new Beckman Coulter DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Abbott Diagnostics Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) and Beckman Coulter LH 780 hematology analyzers using 430 adult specimens. The DxH 800 provided a CBC and differential that correlated well with those of the Sapphire and LH 780, with most parameters showing correlation coefficients (r) of more than 0.97. In the instrument vs 400-cell manual differential comparison, all 3 instruments showed similar and acceptable accuracy to the reference method except for nucleated RBC (NRBC) enumeration, in which the DxH 800 and Sapphire outperformed the LH 780. We also compared clinical efficiency by determining whether flagged specimens showed abnormalities on a peripheral blood smear as defined by International Council for Standardization in Haematology criteria. The efficiency, sensitivity, and specificity of the DxH 800 were 77.0%, 87.1%, and 73.0%, respectively, compared with the Sapphire at 75.8%, 93.5%, and 68.8%, respectively, and LH 780 at 66.1%, 93.5%, and 55.3%, respectively.
We evaluated the new UniCel DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) using 156 pediatric specimens in Microtainer tubes (Becton Dickinson, Franklin Lakes, NJ). The CBC and differential showed good interinstrument correlation, including WBCs (r = 0.995), RBCs (r = 0.992), hemoglobin (r = 0.998), mean corpuscular volume (r = 0.988), platelets (r = 0.997), neutrophils (r = 0.988), lymphocytes (r = 0.984), monocytes (r = 0.815), eosinophils (r = 0.840), basophils (r = 0.049), and nucleated RBCs (NRBCs; r = 0.906). In the instrument vs 400-cell manual differential comparison, the DxH 800 and Sapphire showed comparable performance for nearly all parameters except for NRBCs, for which the DxH 800 correlated better (r = 0.989) than the Sapphire (r = 0.906). We also compared clinical efficiency by determining whether flagged specimens showed abnormalities on a peripheral blood smear as defined by International Council for Standardization in Haematology criteria. The efficiency of the DxH 800 was 78.0% vs the Sapphire at 68.1%. Both instruments showed identical sensitivity (91.1%), but the specificity for the DxH 800 (71.9%) was higher than that of the Sapphire (57.3%).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.