We report the case of a 67-year-old woman who was admitted for surgical removal of a Greenfield filter that had been inserted 7 years before because of recurrent deep vein thrombosis associated with pulmonary embolism. This complication appeared on a plain abdominal radiogram that showed a 7 cm distal migration of the filter, a 30-degree angulation, and rupture of a strut at the level of the hub. Computed tomography, aortography, and ascending cavography demonstrated that the inferior vena cava was perforated by the struts and that the ruptured strut had penetrated the infrarenal aorta. As demonstrated by scanning electron microscopy, the fracture was due to a structural defect of the strut at its insertion point within the hub, with no sign of corrosion. Energy-dispersive radiography analysis failed to demonstrate impurity in the metal composition.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.