Full‐wave electromagnetic simulations have been employed to analyze the design approaches that allow for solving the problems preventing inductive performance of micro‐solenoidal RF probes and formation of strong and homogeneous magnetic fields inside probes at the studies of millimeter‐range sized objects in 14 T MRI scanners. In particular, the effects of non‐uniform coil wrapping on field homogeneity inside extended coils and of partitioning the coils by dielectric separators on coil self‐resonances have been investigated. The possibility to utilize tunable C‐C matching circuits with the coils and to mitigate the effects of sample insertion on the probe resonance frequency has been demonstrated. Challenges of coil fabrication have been also addressed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.