Youth baseball is a relatively safe sport for children. Although injury rates and the total number of injuries declined during the study period, our findings indicate that there are opportunities for making baseball an even safer sport for children. We recommend that all youth baseball players wear properly fitted mouth guards, that all leagues, schools, and parks install safety bases, that all batters use helmets with face shields, and that all players use safety baseballs.
Objectives/Hypothesis
Intraoperative neural monitoring is a useful adjunct for the laryngeal nerve function assessment during thyroid and parathyroid surgery. Typically, monitoring is performed by measurement of electromyographic responses recorded by endotracheal tube (ETT) surface electrodes. Tube position alterations during surgery can cause displacement of the electrodes relative to the vocal cords, leading to false positive loss of signal. Numerous reports have denoted monitoring equipment–related issues, especially endotracheal tube displacement, as the dominant source of false positive error. The false positive error may result in inappropriate decisions by the surgeon. This study tests the hypothesis that anterior laryngeal electrodes (ALEs) can help reduce this error. Placement of ALEs directly onto the thyroid cartilage represent an adjunctive and possible alternative method to standard ETT surface electrodes.
Study Design
Retrospective review.
Methods
Fifteen consecutive patients undergoing thyroid and parathyroid surgery with intraoperative neuromonitoring using both ETT electrodes and ALEs were studied. Data collected included site of neural stimulation, laterality, and electromyographic parameters.
Results
With vagal and recurrent laryngeal nerve stimulation, the ALEs recorded mean vocalis muscle waveform amplitude within 83% of that recorded with standard ETT electrodes. The latency measurements with the anterior laryngeal and endotracheal electrodes were similar, with both electrodes recording significantly longer latency for the left vagus nerve as compared to the right vagus nerve. With superior laryngeal nerve stimulation, the ALEs recorded a 800% greater mean amplitude than the ETT electrodes. The ALEs demonstrated similar sensitivity to stimulation at low current as ETT electrodes and provided stable intraoperative monitoring information.
Conclusions
Compared to ETT surface electrodes, the ALEs provide similar and stable electromyographic responses with equal sensitivity for recording evoked responses during neural monitoring in thyroid and parathyroid surgery. The ALEs offer significantly more robust monitoring of the external branch of the superior laryngeal nerve. Furthermore, ALEs are contained within the operative field, are totally surgeon controlled, and are unaffected by the potential vicissitudes of ETT position during surgery.
Level of Evidence
4 Laryngoscope, 128:2910–2915, 2018
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.