Pediatric cardiac catheterization procedures have the potential to transmit high X-ray doses, which may lead to acute effects or latent skin reactions. Direct measurement of radiation dose was facilitated using nanodot dosimeters and radiochromic film. Direct measurement results were compared with vendor-listed dosimetry and calculation using phantom data. Vendor-listed data demonstrated a wide discrepancy with measured data, whereas the calculation reproducibly overestimated the actual dose. A simple formula was derived to calculate the dose using fluoroscopy time, cine frame quantity and average cine mA in a biplane environment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.