We describe a multinational evaluation of the Menarini– Arkray HA 8140 hemoglobin (Hb) A1canalyzer, which utilizes a high degree of automation, including bar code reading, cap piercing, and whole-blood sampling. Within- and between-batch CVs were <2%. Linearity was confirmed throughout the working range of the analyzer. Common Hb variants, including Hb S, Hb C, and Hb F, did not interfere with the Hb A1c separation, and the potentially interfering labile Schiff base was effectively removed during the chromatographic procedure. The HA 8140 analyzer displayed good correlation to the Bio-Rad Variant analyzer, Tinaquant immunoassay, affinity chromatography, and an optimized “in-house” HPLC Hb A1c method. The methods when compared by Altman and Bland plots showed bias (upper, lower 95% confidence limits) of: Variant minus HA 8140 = 0.99 (0.23, 1.74), Tinaquant minus HA 8140 = 0.14 (−0.71, 0.98); affinity minus HA 8140 (after log transformation) = 1.13 (0.90, 1.41), and “in house” HPLC minus HA 8140 (after log transformation) = 0.91 (0.82, 1.01).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.