Methods have been developed for using the Beckman "Discrete Sample Analyzer" (DSA-560), for automated simultaneous determinations of p-aminohippurate (PAH) and creatinine in the same 50- or 10-µl samples of plasma or urine, respectively, at the rate of 80 samples per hour. In determinations of PAH, a single reagent, p-dimethylaminocinnamaldehyde in 0.1 mol/liter HCl, was added to a protein-free filtrate from plasma or urine. The intensity of the color, measured at 550 nm, obeyed Beer’s law for PAH concentrations from 0.005 to 5.0 g/liter. Data obtained for the same plasma and urine samples by the manual method, were, in general, within ±5% of the results obtained by the automated method. Creatinine was simultaneously determined by adapting the method of Taussky to the DSA-560 instrument; the intensity of the color, measured at 510 nm, obeyed Beer's law for creatinine concentrations ranging from 0.005 to 5.0 g/liter. Data obtained for the same plasma and urine samples by the manual method, were, in general, within ± 5% of the results obtained by the automated method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.