The factor V Leiden mutation, a G→A transition at position 1691 in exon 10 of the gene that codes for factor V, produces an Arg506Gln substitution and is the most common genetic risk factor for venous thrombosis. We have developed a rapid, sensitive, and specific method to detect the factor V Leiden mutation in genomic DNA from whole blood by PCR amplification and microparticle enzyme immunoassay detection using the Abbott LCx instrument. We compared this automated method with the standard procedure using restriction endonuclease digestion of PCR products followed by gel electrophoresis in blinded experiments. In 130 patients (from Veterans Affairs medical centers) with deep venous thromboses, including 24 heterozygotes with the factor V Leiden mutation, there was complete agreement between the two methods. The assay was also able to distinguish heterozygotes from homozygotes. This method, which carries a low potential for cross-contamination of samples, should be a useful routine test for the factor V Leiden mutation in clinical laboratories with sufficient demand for molecular diagnostic assays using the LCx instrument.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.