Objectives
Hybrid closed loop systems (HCL) improve the management of type 1 diabetes (T1DM). T1DM adolescent patients represent a risk category also if they are in an automated insulin infusion delivery therapy.
Case presentation
We describe a series of four cases in which adolescent patients have adopted incorrect behaviours in the managing of HCL systems, challenging the algorithm skills. Two patients performed fabricated sensor calibrations. The other two did not perform pre-prandial insulin boluses correctly. Despite these behaviours, the algorithm corrected the glucose values in three out of four patients. Only in one case, where fabricated calibrations were too frequent, the automatic system failed to restore the glycemic balance.
Conclusions
Fabricated calibrations seem to be more important than uncorrected insulin boluses to challenge the HCL systems.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.