A simple, rapid, and sensitive assay of nicotine in tobacco has been developed using pyrolysis-gas chromatography-mass spectrometry (DSP-GC-MS). The optimum pyrolyzer (desorption) conditions, using Korean tobacco grade B1O (0.4 mg) and n-heptadecane as an internal standard, were found to be heating from 50 to 300 degrees C at a rate of 60 degrees C/min. Replicate determinations (n = 5), on the same tobacco and using n-heptadecane as the internal standard, resulted in a nicotine content of 1.96%, with a relative standard deviation (RSD) of 1.9%. This result was in good agreement with those from established methods: the Cai ether extraction method, a chloroform extraction method, and the CORESTA recommended method. However, the DSP method requires less than half the time of the solvent extraction methods, requires less sample, is almost solvent-free, and is less labor-intensive. The DSP method was used to determine the nicotine content of eight flue-cured tobaccos from Brazil, China, Korea, and the United States, which were found to have contents ranging from 1.21 to 2.19%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.