Although the analytical literature seems abundant for the determination of metoprolol in human plasma, a method using standard equipment providing a sensitive and simple high-performance liquid chromatographic (HPLC) method for limited blood volume, e.g. where 1 mL of blood in a 1 kg infant equals 70 mL of adult blood volume, has rarely been addressed. Therefore, in 500 microL of plasma, metoprolol was extracted using an internal standard and solid-phase extraction columns. Chromatographic analysis was performed on a Spherisorb C(6) column (5 microm particle size) at ambient temperature and fluorimetric detection with an excitation wavelength of 225 nm, and emission wavelength of 310 nm. The mobile phase [30% acetonitrile and 70% 0.25 m potassium acetate buffer (pH 4)] was pumped with 1 mL/min. Metoprolol recovery was determined at 73.0 +/- 20.5%, and the limit of quantitation was 2.4 ng/mL. Precision values of intra- and inter-assay were below 15.5% and those for accuracy were between 90 and 110%. This method was developed for monitoring and determination of pharmacokinetic parameters of metoprolol in pediatric patients and therefore metoprolol plasma concentrations in a 2-year-old child with ventricular tachycardia are reported. .
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.