Patients with THA using a ceramic head and using an OxZr head were each matched to patients with a CoCr head. Mean implantation time was 5 years. There was no difference in steady state wear rate between the ceramic (0.066 ± 0.050 mm/year) and CoCr match groups (0.052 ± 0.041 mm/year), or between the OxZr (0.022 ± 0.029 mm/year) and CoCr match groups (0.048 ± 0.071 mm/year). Follow-up into the second decade will be necessary before any changes in THA wear rate from using ceramic or OxZr bearings may be appreciated clinically with available imaging techniques.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.