NEMA characterization of PET systems is generally based on 18 F because it is the most relevant radioisotope for the clinical use of PET. 18 F has a half-life of 109.7 min and decays into stable 18 O via β+ emission with a probability of over 96% and a maximum positron energy of 0.633 MeV. Other commercially available PET radioisotopes, such as 82 Rb and 68 Ga have more complex decay schemes with a variety of prompt gammas, which can directly fall into the energy window and induce false coincidence detections by the PET scanner. Methods: Aim of this work was threefold: (A) Develop a GATE model of the GE Signa PET/MR to perform realistic and relevant Monte Carlo simulations (B) Validate this model with published sensitivity and Noise Equivalent Count Rate (NECR) data for 18 F and 68 Ga (C) Use the validated GATE-model to predict the system performance for other PET isotopes including 11 C, 15 O, 13 N, 82 Rb, and 68 Ga and to evaluate the effect of a 3T magnetic field on the positron range. Results: Simulated sensitivity and NECR tests performed with the GATE-model for different radioisotopes were in line with literature values. Simulated sensitivities for 18 F and 68 Ga were 21.2 and 19.0 /kBq, respectively, for the center position and 21.1 and 19.0 cps/kBq, respectively, for the 10 cm off-center position compared to the corresponding measured values of 21.8 and 20.0 cps/kBq for the center position and 21.1 and 19.6 cps/kBq for the 10 cm off-center position. In terms of NECR, the simulated peak NECR was 216.8 kcps at 17.40 kBq/ml for 18 F and 207.1 kcps at 20.10 kBq/ml for 68 Ga compared to the measured peak NECR of 216.8 kcps at 18.60 kBq/ml and 205.6 kcps at 20.40 kBq/ml for 18 F and 68 Ga, respectively. For 11 C, 13 N, and 15 O, results confirmed a peak NECR similar to 18 F with the effective activity concentration scaled by the inverse of the positron fraction. For 82 Rb, and 68 Ga, the peak NECR was lower than for 18 F while the corresponding activity concentrations were higher. For the higher energy positron emitters, the positron range was confirmed to be
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.