The main goal of this work was development of ECG processing algorithm for the real-time detection and displaying of implanted pacemaker artifacts, if they are present in the signal. In case of traditionally used characteristics of ECG acquisition circuits and analogto-digital conversion pacemaker artifacts are often scarcely distinguished. Wide frequency band of ECG amplifier (0.2-10000 Hz) and high ADC sampling rate (20000 Hz) were used to resolve this problem. All pacemaker spikes detection procedures were realized fully digitally. The developed algorithm testing with the use of the real ECG recordings set showed the following results of pacemaker spikes detection: sensitivity 86.6 % and positive predictivity 99.6 %. The proposed algorithm was included in the ECG processing software of mobile cardiac monitor CardioQVARK based on iPhone series 5, 5S and 6.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.