Proton Fourier transform nuclear magnetic resonance spectral data were collected on n-alcohol ethoxylates and used to calculate the percent ethylene oxide (EO) content of the n-alcohol ethoxylate without an internal or external standard. The accuracy and precision of the method were determined from ten repetitive analyses of hexaethylene glycol mono n-dodecyl ether. The standard deviation was 0.23 wt% EO with a relative standard deviation of 0.40%. The method had a relative error of +0.55% and an absolute error of +0.32% EO.KEY WORDS: Ethylene oxide, n-alcohol ethoxylatc analysis, NMR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.