Aims
Sudden cardiac death (SCD) is the most common mode of death in paediatric hypertrophic cardiomyopathy (HCM). This study describes the implant and programming strategies with clinical outcomes following implantable cardioverter-defibrillator (ICD) insertion in a well-characterized national paediatric HCM cohort.
Methods and results
Data from 90 patients undergoing ICD insertion at a median age 13 (±3.5) for primary (n = 67, 74%) or secondary prevention (n = 23, 26%) were collected from a retrospective, longitudinal multi-centre cohort of children (<16 years) with HCM from the UK. Seventy-six (84%) had an endovascular system [14 (18%) dual coil], 3 (3%) epicardial, and 11 (12%) subcutaneous system. Defibrillation threshold (DFT) testing was performed at implant in 68 (76%). Inadequate DFT in four led to implant adjustment in three patients. Over a median follow-up of 54 months (interquartile range 28–111), 25 (28%) patients had 53 appropriate therapies [ICD shock n = 45, anti-tachycardia pacing (ATP) n = 8], incidence rate 4.7 per 100 patient years (95% CI 2.9–7.6). Eight inappropriate therapies occurred in 7 (8%) patients (ICD shock n = 4, ATP n = 4), incidence rate 1.1/100 patient years (95% CI 0.4–2.5). Three patients (3%) died following arrhythmic events, despite a functioning device. Other device complications were seen in 28 patients (31%), including lead-related complications (n = 15) and infection (n = 10). No clinical, device, or programming characteristics predicted time to inappropriate therapy or lead complication.
Conclusion
In a large national cohort of paediatric HCM patients with an ICD, device and programming strategies varied widely. No particular strategy was associated with inappropriate therapies, missed/delayed therapies, or lead complications.
Timothy syndrome is a rare disorder with a high attrition rate if undiagnosed. Perioperative cardiac arrests are common and not always amenable to resuscitation. Longer-term survival is possible, however, patients invariably require pacemaker or defibrillator implantation.
We describe the use of a new vascular occlusion device to embolize vessels whose flow is detrimental to cardiac function in a variety of clinical situations in both adults and children with congenital cardiac disease. Our series includes four cases in which we have successfully occluded large venous vessels, to produce a beneficial effect on oxygen saturations and cardiac function. There were no adverse events because of the devices in our series, which provides evidence that the Amplatzer vascular plug is a safe and effective method of embolizing large venous vessels.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.