DETERMINATION OF Cd, Pb AND Zn IN FRESH WATERS, SEAWATER AND MARINE SEDIMENTS. Thermospray flame furnace Atomic Absorption Spectrometry (TS-FF-AAS) was used for the total determination of Cd, Pb and Zn in fresh water and seawater samples at µg L-1 levels, and in marine sediment samples at µg g-1 levels. Using a sample loop of 50 µL and a peristaltic pump the samples were transported into the metallic tube placed over an air/acetylene flame, through a ceramic capillary (o.d. = 3.2 mm) containing two parallel internal orifices (i.d = 0.5 mm). The detection limits determined for Cd, Pb and Zn using a synthetic water matrix (2.5% m/v NaCl, 0.5% m/v MgCl 2 and 0.8% m/v CaCl 2) were 0.32 µg L-1 ; 2.6 µg L-1 and 0.21 µg L-1 respectively. The methodology by TS-FF-AAS was validated by determination of Cd, Pb and Zn in certified reference materials of water and marine sediment, and the t-test for differences between means was applied. No statistically significant differences were established in fresh water and seawater (p>0.05), whereas differences became apparent in marine sediment (p<0.03).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.