In this study population of procedurally naive medical students, telepresent supervision was as effective as in-person supervision for tracheal intubation.
SummaryA patient received a massive overdose of papaveretum intravenously (estimated to be 180 mg) when the glass syringe of a patient-controlled analgesia machine disengaged from the drive mechanism. She was successfully resuscitated. The pump, on loan ,from the supplier, had passed a brief evaluation by the infusion pump test house designated by the Medical Devices Directorate qf the Department of Health; it has since been withdrawn. It is recommended that patient-controlled analgesia equipment should be placed at or below patient heart level. The Department of Health is called on to institute a full. independent evaluation scheme for patient-controlled analgesia equipment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.