We measured temperature on admission to the neonatal unit in a cohort of 54 very preterm infants. We measured rectal temperature with a digital thermometer (Microlife MT-1931) as the gold standard (MT-R). We also measured axillary temperature with the MT (MT-A), with the Welch Allyn SureTemp Plus 692 in ‘continuous’ (WAC) mode and in the default ‘predictive’ (WAP) mode. While MT-A and WAC frequently differed from MT-R by ≥0.3°C, they were both reasonably sensitive and specific for hypothermia (MT-R <36.5°C). WAP overestimated MT-R by ≥0.5°C on 37/53 (70%) occasions and had poor sensitivity for hypothermia, identifying only 2 of 29 infants with MT-R <36.5°C as hypothermic.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.