We developed a modified nitrogen washin/washout technique based on standard monitors using inspiratory and end-tidal gas concentration values for functional residual capacity (FRC) measurements in patients with acute respiratory failure (ARF). For validation we used an oxygen-consuming lung model ventilated with an inspiratory oxygen fraction (Fio(2)) between 0.3 and 1.0. The respiratory quotient of the lung model was varied between 0.7 and 1.0. Measurements were performed changing Fio(2) with fractions of 0.1, 0.2, and 0.3. In 28 patients with ARF, duplicate measurements were performed. In the lung model, an Fio(2) change of 0.1 resulted in a value of 103 +/- 5% of the reference FRC value of the lung model, and the precision was equally good up to an Fio(2) of 1.0 with a value of 103 +/- 7%. In the patients, duplicate measurements showed a bias of -5 mL with a 95% confidence interval [-38; 29 mL ]. A comparison of a change in Fio(2) of 0.1 with 0.3 showed a bias of -9 mL and limits of agreement of [-365; 347 mL]. This study shows good precision of FRC measurements with standard monitors using a change in Fio(2) of only 0.1. Measurements can be performed with equal precision up to an Fio(2) of 1.0.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.