A flow-through optosensor for anthracyclines based on the anthracycline-europium chelate room temperature phosphorescence energy transfer is proposed. The sensor was developed in conjunction with a flow injection analysis system and is based on the transient immobilization on a non-ionic resin (packed in a flow-through cell) of the anthracycline-europium chelate. The analytical performance characteristics of the proposed sensor for semi-automated analysis and control of very low levels of anthracycline were as follows: the detection limits for daunorubicin, doxorubicin and epirubicin were 9.0, 5.8 and 5.8 ng ml 21 , respectively, with an RSD of 1% for the determination of 0.22 mg ml 21 of each anthracycline (n = 10). Most of the common metal ions in biological samples did not interfere, except Fe III , which caused serious interference and should be masked with 1,10-phenanthroline. The recommended method was successfully tested for determination of anthracyclines in clinical samples (urine and pharmaceutical preparations).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.