Continuous flow delivery of nitric oxide into the circuit of a phasic-flow ventilator results in marked inspiratory nitric oxide concentration fluctuation that is not detected by a slow-response chemiluminescence analyzer. Moreover, nitric oxide concentration fluctuation can influence the accuracy of the chemiluminescence measurements. These effects can be diminished by using additional mixing chambers to facilitate a stable gas concentration. As these mixing volumes increase the contact time of nitric oxide with oxygen, an increase of nitrogen dioxide has to be taken into account.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.