An analysis of electrocardiographic pattern recognition parameters for premature ventricular contraction (PVC) and normal (N) beat classification is presented. Twenty-six parameters were defined: 11 x 2 for the two electrocardiogram (ECG) leads, width of the complex and three parameters derived from a single-plane vectorcardiogram (VCG). Some of the parameters include amplitudes of maximal positive and maximal negative peaks, area of absolute values, area of positive values, area of negative values, number of samples with 70% higher amplitude than that of the highest peak, amplitude and angle of the QRS vector in a VCG plane. They were measured for all heartbeats annotated as N or PVC in all 48 ECG recordings of the MIT-BIH arrhythmia database. Two reference sets for the Kth nearest-neighbours rule were used-global and local. The classification indices obtained with the global reference set were 75.4% specificity and 80.9% sensitivity. Using the local reference set we increased the specificity to 96.7% and the sensitivity to 96.9%. The achieved specificity and sensitivity are comparable with, and greater than, the results reported in the literature.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.