Using an innovative time-varying front-end electronics in conjunction with a bulky coaxial high-purity germanium detector, we were able to extend the range of the radionuclide spectra well beyond the analog to digital converter (ADC) saturation point. The electronics automatically conditions the signal for digital-filtering optimization if it is in the ADC voltage range and for time-over-threshold analysis if it exceeds the ADC range. A high spectroscopic resolution has been achieved in both operation ranges. An unprecedented wide energy range from 5 keV to 150 MeV of equivalent energy, or 90 dB, has been obtained using a single acquisition chain, while maintaining a high-energy resolution in the whole spectrum. For example, with an ADC range of 3 MeV a resolution has been obtained of 1.3/2.2 keV full width at half maximum on the 122/1332 keV gamma-ray lines of 57Co and 60Co, and of <0.4% in the time-over-threshold region, or for energy deposits beyond 3 MeV.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.