The reliability of ACM is high over a long follow-up. On the basis of atrial threshold variability, a practical approach to ACM programming should be two daily atrial threshold measurements in patients with normal AV conduction and%Ap < or = 40%, or with normal sinus activity and AVB, whereas 3-4 measurements should be recommended in patients with first-degree AVB and%Ap > 40% or with recurrent atrial fibrillation and AVB. The lowest adapted stimulation output should achieve at least threshold +1 V in the first trimester after implantation, then threshold +0.5 V thereafter, in order to achieve > 99.5% effective atrial stimulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.