Background Acetabular screw cups seem to give high primary stability. We analyzed the migration and loosening behavior of a first-generation screw cup in a longterm follow-up.Patients and methods We examined 92 uncemented titanium alloy conical screw cups prospectively. Implant migration was assessed with a digital high-precision method (EBRA) with an accuracy of 1.0 mm.Results After mean 11 (0.5-18) years, 87 patients were available for examination and 5 patients had died. 32 implants had been revised and 7 cases showed radiographic evidence of loosening. The 10-year survival rate was 71%. Migration of more than 1 mm occurred in 53 hips. Implant survival was strongly associated with an annual migration of greater than 0.2 mm.Interpretation The long-term behavior of this cup is not satisfactory. In spite of extraordinarily high primary implant stability, secondary osseointegration of this cup often fails. The annual migration rate represents a valid parameter for prediction of implant survival.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.