SummaryThe purpose of this study was to develop an optimised sample preparation method for the determination of Pu, Am, and Cm in environmental samples, soils, and sediments, and to compare the sensitivities of DF-ICP-MS, alpha spectrometry, and liquid scintillation alpha spectrometry. Borate fusion was used for sample material digestion. Two methods based on TOPO liquid–liquid extraction and combination of TEVA and TRU extraction chromatography showed highest yields of 92.7% for Pu, 86.8% for Am, and 87.6% for Cm. The detection limits were between 0.014–0.034 Bq/kg for
Neptunium / Inductively coupled plasma mass spectrometry / Alpha spectrometry / Extraction chromatography / Soils Summary. A fast and simple method for the determination of 237 Np in soil is presented. The borate fusion decreases the sample pre-treatment time and a TEVA extraction chromatography separates neptunium from uranium and the interfering matrix components. A comparison of the sensitivities of alpha spectrometry and sector field ICP-MS for determination of 237 Np is presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.