The effects of reducing the pulse repetition time from 2500 ms to 1000 ms when using spin-density-projection-assisted R2-magnetic resonance imaging for the purpose of measuring liver iron concentration were evaluated. Repeated liver R2 measurements were made using both protocols on 60 subjects with liver iron concentrations ranging from 0.5 to 48.6 mg Fe (g dry tissue)21 . The mean total scan time at repetition time 1000 ms was 42% of that at repetition time 2500 ms. The repeatability coefficients for the two protocols were not significantly different from each other. A systematic difference in the measured R2 using each protocol was found indicating that an adjustment factor is required when one protocol is used to replace the other. The 95% limits of agreement between the two protocols were not significantly different from their repeatability coefficients indicating that the protocols can be interchanged without any significant change in accuracy or precision of liver iron concentration measurement. Magn Reson Med 65:1346-1351, 2011. V C 2010 Wiley-Liss, Inc.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.