Some primary concerns with freehand 3D ultrasound are time-consuming calibration operations, overall accuracy, and computation time during the 3D reconstruction process. A procedure was developed to reduce the calibration time to only 10 seconds by planning the path of the probe. A decrease in marginal error may improve the overall accuracy by searching axial and lateral distances with a maximum correlation coefficient, and was based on constructing an adaptive decorrelation curve from images containing Rayleigh and coherent scatters. Marginal errors were reduced in the range of 3% to 11% when the distance estimation was between 0.15mm and 0.375mm in in-vitro studies. To save computation time within the 3D reconstruction, a low-resolution 3D reconstruction algorithm is proposed so that verification can be completed prior to computing a high-resolution reconstruction.ii
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.