Abstract:Calibration of an electric field probe using a different type and size of probe as a reference is carried out in a GTEM cell from 1 to 6 GHz. We compared our results with those of calibration in an anechoic chamber to determine the effectiveness of the calibration. The difference between the calibration factors of the GTEM cell and the anechoic chamber increases when the probe is calibrated in an area in which the electric field distribution is nonuniform. Therefore, it is important to place the probe in an area with good uniformity or maintain the uniformity of the electric field in the GTEM cell when calibrating electric field probes with different dimensions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.