We have designed and demonstrated a standoff Raman system for detecting high explosive materials at distances up to 50 meters in ambient light conditions. In the system, light is collected using an 8-in. Schmidt-Cassegrain telescope fiber-coupled to an f/1.8 spectrograph with a gated intensified charge-coupled device (ICCD) detector. A frequency-doubled Nd : YAG (532 nm) pulsed (10 Hz) laser is used as the excitation source for measuring remote spectra of samples containing up to 8% explosive materials. The explosives RDX, TNT, and PETN as well as nitrate- and chlorate-containing materials were used to evaluate the performance of the system with samples placed at distances of 27 and 50 meters. Laser power studies were performed to determine the effects of laser heating and photodegradation on the samples. Raman signal levels were found to increase linearly with increasing laser energy up to approximately 3 x 10(6) W/cm2 for all samples except TNT, which showed some evidence of photo- or thermal degradation at higher laser power densities. Detector gate width studies showed that Raman spectra could be acquired in high levels of ambient light using a 10 microsecond gate width.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.