This case report describes delivery of atrial therapies during a sinus tachycardia in a new dual chamber implantable cardioverter defibrillator inappropriately caused by far-field oversensing of ventricular beats in the atrial channel. Upon classification of the PR interval pattern, the rate criterion for an atrial tachycardia was fulfilled, and the device initiated high-frequency burst pacing as the first stage of programmed tiered atrial therapies. Atrial fibrillation subsequently was induced by high-frequency burst pacing, and eventually a programmed 10-J shock was delivered for successful termination of atrial fibrillation. The phenomenon of far-field oversensing of ventricular beats could be repeatedly observed during exercise testing and abolished by decreasing the atrial sensitivity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.