Several voltammetric techniques were used to explore the reductive behaviour of the antihypertensive agent doxazosin on a bare carbon paste (CPE) and a Tenax-modified carbon paste electrode (TMCPE). The results indicate that the process is irreversible and fundamentally controlled by adsorption, which allows doxazosin to be accumulated at the electrode surface. The cathodic adsorptive stripping (AdS) response was evaluated with respect to pH, accumulation variables and instrumental parameters, using differential-pulse (DPV) and square-wave voltammetry (SWV) as redissolution techniques. In both cases, a voltammetric peak close to 0 V in Britton-Robinson buffer (pH 6.6) was obtained after a preconcentration step at 0.55 V for 3 min (2000 rpm) and a subsequent cathodic scan. When the TMCPE was used, the limits of detection were 4.35 x 10(-11) and 5.18 x 10(-11) M for AdS-DPV and AdS-SWV, respectively. The deposition time (3 min) was improved relative to that obtained by means of CPE (6 min). Under the optimum operational conditions, the doxazosin reduction peak showed a linear response in the range from 6 x 10(-11) to 1 x 10(-9) M by using AdS-DPV, with an RSD of 3.39% for 3 x 10(-8) M doxazosin solution (n = 10). A method was developed for the determination of doxazosin in human urine and formulations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.