The new facility Artemis was installed in 2003 in Saclay, France. This 3MV NEC Pelletron is dedicated to high-precision radiocarbon measurements for French 14C laboratories. We will present information on Artemis along with our sample preparation methods. Results from measurements on some intercalibration samples will be given along with the values of measured blanks. Finally, we report on some problems we have encountered when measuring sputter cathodes with high CH− outputs.
The main objective of this report is to present the dating process routinely applied to different types of samples at the Laboratoire de Mesure du Carbone 14 (LMC14). All the results and protocols refer to our procedures over the last 5 years. A description of the sorting and chemical pretreatments of the samples as well as the extraction and graphitization of CO2 are reported. Our last study concerning the degradation of the blank level according to the storage time of the targets between graphitization and accelerator mass spectrometry (AMS) measurement is also presented. This article also provides information on how to submit a valid laboratory sample. We give details relating to sampling procedures on site as well as contamination issues relative to the 14C dating methodology.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.